Medical packaging and conditioning in cleanroom
Certificated ISO 13485
ISO 13485 certificate

ULTRASON BATH DETERGENT MECA WASH

ULTRASON BATH DETERGENT
Detergent solution dedicated to ultrasonic cleaning and disinfection of medical implants.

The wash water is dedicated to ultrasonic cleaning and disinfection of medical implants. Equipped with a brushless motor 2 robot axes very precise, this line of ultrasonic cleaning is suitable for all types of your medical devices. All tanks are interchangeable and can be configured to suit your needs. Two types are possible oven (atmospheric and vacuum oven thermal oven) depending on the complexity of your DM.

Description bath detergent :

  • 1 era tank (laundry) : 500 x 200 x 480, ultrasound (40 KHz, 640W), thermo regulated (1000W).
  • 2 nd tank (rinse) : 500 x 200 x 480.
  • 3 rd tank (laundry) : 300 x 200 x 480, 3 rd bowl (laundry) (40 KHz, 640W), thermo regulated (1000W).
  • 4 th tank (rinse) : 500 x 200 x 480.
  • 5 th tank (rinse) : 500 x 200 x 480.
  • 4 th tank (rinse purified water) : 500 x 200 x 480, ultra-sons (40 KHz, 640W), thermo régulée (1500W).
  • Drying oven vacuum : 500 x 200 x 600, thermo regulated (1300W).
  • Dimensions 3000 x 780 x h 2400 mm.
  • 2-axis robot for handling baskets.
  • Touch screen control and control via PLC.
  • Enregistrement des paramétres de production sur clé USB.
  • Outputs provided for easy calibration.

Features bath detergent :

Accuracy of various parameters

  • Frequency US : +/- 1KHz.
  • Heater : +/- 3,0°C.
  • Time : +/- 1 second.
  • Depression : +/- 10mbar.

Limitations of use

  • Frequency US : 40 KHz.
  • Power US : 0 to 640 W
  • heated tanks : 25°C to 60°C.
  • heated oven : 25°C to 110°C.
  • Depression : 20 mbar to 1 bar.

Energies required

  • Compressed air : 7 bars.
  • Electricity : 400V three-phase

General

  • ultrasonic generator capable of tracing the frequency and power.
  • Robot 2 axes making handling baskets.
  • Installation of low-rise baskets by operators minimizing physical effort.
  • 4 controllers (PID) heating with digital communication between the controller and the controller, allowing no loss of information related to temperature.
  • Time scheduling of the stills to a rise in temperature and a temperature stabilization before starting production.
  • Traceability of each cycle on USB (date, time, frequency US, US power, time, temperature, conductivity of the final rinse water) key.
  • Calibration mode (time, temperature and pressure).
  • Folder QC, IQ, OQ, PQ available on request at extra cost.
  • Compliance with safety.
  • The pneumatic part is made with material SMC.
  • The controller, display and bruschless engines Schneider brand.
  • Temperature control is performed with hardware Geffran.